Peak Resolution
One critical specification for gamma ray detectors is peak
resolution. In nuclear spectroscopy, we often express the
resolution as a percentage at 662 keV (cesium). For the SAM
Revealer, the typical resolution is 2.8%. This is about 2.5
times better than sodium iodide detectors and therefore
allows separation of many closely spaced peaks.
The Lanthanum Bromide Advantage
Because your applications for isotope identification vary,
and user skill level varies, we offer several modes of
operation. Training requirements are reduced by the handy
user interface and color-coded isotope categories. A SAM
Revealer Easy- Mode simplifies screening and identification
activities. In cases where you are identifying special
nuclear material or masked isotopes, enhanced peak
separation as found in the SAM Revealer will be helpful in
giving you high confidence reporting. For example,
separating the conflicting and overlapping lines of multiple
sources such as Cs137 and Th232, Cs137 and Ra226 or Co60 and
Ra226, is challenging with low resolution detectors (6% or
greater). Advanced algorithms can compensate for some of the
shortfall, but high resolution from a LaBr detector
identification with a high degree of confidence. One of the
best advantages of LaBr is clearly seeing the gamma and x-
ray peaks of plutonium. The one highly penetrating NaI peak
of Pu239 is now clearly seperated into three easily
recognized peaks with LaBr . The first line responder can
easily email the SAM Revealer data report (complete with GPS
data, ANSI N42.42 compliant) to the appropriate scientific
support center for further analysis. These enhanced spectra
will be of great value to those responsible for evaluating
and responding to critical situations.